Study Design:Prospective,randomized,clinical trial on 200 patients randomized to groups 1(n=80)Centurion vision system with traditional handpiece, group 2(n=80)Centurion vision system with active sentry handpiece and group 3 Infiniti vision system. Within groups 1&2,there were 2 subgroups where low(30mmHg) and high IOP(50mmHG)were used.Realtime dynamic intraoperative IOP was measured.Cumulative dissipated energy(CDE),% change in central corneal thickness(CCT),corneal endothelial morphology at 6 months postoperatively
Results: CDE was significantly lower with Active Sentry(4.14±2.25vs2.96±1.4, p=0.001). Within traditional handpiece,low IOP showed lesser %increase in CCT on day 1(0.09 ± 0.1Vs 0.15 ± 0.1,p=0.01)and week 1 compared to high IOP(0.02±0.02 vs0.03±0.02,p=0.03). % change in endothelial cells was significantly lower when using active sentry handpiece (-0.957+0.17vs-0.98+0.004,p=0.005)
Conclusion: Active sentry performed well at varying IOP


FP2006 : INTRAOPERATIVE PERFORMANCE AND OUTCOMES DURING PHACOEMULSIFICATION WITH THREE FLUIDIC SYSTEMS